Maxum II Specifications, Continued
Gas Sample Requirements
Sample Flow
50-200 cc/min (application dependent)
Sample Filtration 0.1 micron
Minimum Sample Pressure 5 psig (35 kPa), lower pressure optional
Maximum Sample Pressure 75 psig (517 kPa), standard; higher pressure optional for
airbath/airless oven
Maximum Sample Temperature 250°F (121°C) standard for airbath/airless oven; higher temperature
optional
176°F (80°C) for modular oven
(Application dependent: not higher than oven temperature)
Material in Contact with Sample Stainless steel, Teflon©, and polyamide; other material optional
Liquid Sample Requirements
Sample Flow
5-20 cc/min (application dependent)
Sample Filtration 0.3-5 micron (sample valve dependent)
Minimum Sample Pressure 5 psig (35 kPa), lower pressure optional
Maximum Sample Pressure 300 psig (2070 kPa) standard; higher pressure optional
Maximum Sample Temperature 250°F (121°C) standard; higher temperature optional
Material in Contact with Sample Stainless steel and Teflon©, other material optional